EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>G7121B

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
132 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
5
Optimizing the Detector
Selecting the Best Response Time
Selecting the Best Response Time
Selecting the Best Response Time
Data reduction using the RESPONSETIME function will increase your
signal-to-noise ratio.
For example, see Figure 49 on page 132.
Figure 49 Finding Best Response Time
LC fluorescence detectors typically work with response times of 2 or 4 s. The
default of the module is 4 seconds. It is important to know that comparing
sensitivity requires using the same response time. A response time of 4 s
(default) is equivalent to a time constant of 1.8 s and appropriate for standard
chromatographic conditions.
1 sec
VHF
VHF

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals