EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>G7121B

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 171
Test Functions and Calibration
8
Lamp Intensity Test
Lamp Intensity Test
The intensity test scans an intensity spectrum via the reference diode (200 –
1200 nm in 1 nm steps) and stores it in a diagnosis buffer. The scan is
displayed in a graphic window. There is no further evaluation of the test.
Results of this test are stored as lamp history (date code, intensity).
Figure 57 Lamp Intensity Test (Agilent Lab Advisor)
UV degradation, especially below 250 nm is significantly higher compared to
visible wavelength range. Generally the "LAMP ON during run" setting or using
"economy mode" will increase lamp life by a magnitude.
NOTE
The profile can vary from instrument to instrument. It is dependig on the age of the lamp
and the content of the flow cell (use fresh water).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals