158 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
7
Error Information
Detector Error Messages
Flash Lamp Current Overflow
Error ID: 6704
The lamp current of the xenon flash lamp is monitored constantly. If the
current gets too high, an error is generated and the lamp is turned OFF.
No light at reference diode despite lamp is on
Error ID: 6721
Revision A/B/C Front End Board (FLF):
There is no feedback mechanism that checks whether the lamp is ON! If no
peaks are shown in the chromatogram, the user-interface shows the module
still in Ready. Perform a "Lamp Intensity Test" (see “Lamp Intensity Test” on
page 171) first. If flat use below steps.
Revision D Front End Board (FLF):
The flashing of the xenon flash lamp is monitored constantly. If the Lamp
has not flashed for more than 100 times in series, an error is generated and
the lamp is turned OFF.
Probable cause Suggested actions
1
Short-circuit of trigger pack assembly or
defective FLL board.
Please contact your Agilent service
representative.
2
Short-circuit of flash lamp assembly.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defective Hardware.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals