EasyManuals Logo

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
112 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
5
Optimizing the Detector
Example: Optimization for Multiple Compounds
5 Wait until the baseline stabilizes. Start the run.
6 Load the signal.
Figure 41 Chromatogram - Excitation Scan at Reference Wavelength 260/330 nm
7 Use the isoabsorbance plot and evaluate the optimal excitation wavelengths
(in this example just in the time range of 13 minutes).
Figure 42 Isoabsorbance Plot - Excitation
The table below shows the complete information about emission (from
Figure 39 on page 110) and excitation maxima.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals