EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>G7121B

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
164 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
7
Error Information
Detector Error Messages
Motor Speed Unstable
Error ID: EX 6717, EM 6718
For proper operation the monochromator gratings must run at a certain
constant revolution.
Motor Encoder Index Wrong
Error ID: EX 6719, EM 6720
The actual encoder pattern is checked against a known pattern.
Probable cause Suggested actions
1
Defective monochromator assembly.
Re-ignite the lamp.
Replace monochromator assembly.
Probable cause Suggested actions
1
Encoder was replaced and has a different
pattern or no reset of pattern was made.
Reset pattern via user interface and recalibrate.
2
Encoder lost position completely.
Replace monochromator assembly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals