180 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Wavelength Accuracy Test
Figure 64 Example of good EX/EM maxima
If the plots do not have a maximum around EX=397 nm and EX=350 nm
(±3 nm) the test fails, see figure below. Refer to “Interpretation of the
Results” on page 181.