190 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Wavelength Calibration Procedure
4 Verification using the "Wavelength Accuracy Test".
a Refit the capillary to the flow cell.
b Follow the procedure "Wavelength Accuracy Test".
NOTE
Rinse the flow cell with pure water at a minimum of 1.5 mL/min to get rid of the Glycogen
from the cell and the capillaries. When organic solvent is sequentially applied (without
rinsing), a blockage of capillaries may occur.