204 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Other Lab Advisor Functions
FLD Mode
No multi WL: You specify an Excitation wavelength (Ex) and an Emission
wavelength (Em). If you select No multi WL, you cannot set up a scan.
Excitation Scan: You specify an Emission wavelength (Em) and the excitation
scan range.
Emission Scan: You specify an excitation wavelength (Ex) and the emission
scan range.
Scan Parameters:
Flash Lamp On: Switches on the FLD flash lamp.
Sample Scan: Scans the sample spectrum over the specified wavelength
range at the specified resolution and in the selected mode. You specify the
wavelength range in the from and to fields, and the resolution in the step
field. You select the mode in the FLD Mode section.
•Export Data Exports the selected data in csv format for use in other
applications.