EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>G7121B

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #216 background imageLoading...
Page #216 background image
216 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
9
Maintenance
Exchanging a Flow Cell
7
Reconnect the capillaries to the flow cell.
NOTE
DO NOT install the inlet capillary to the outlet
connection of the flow cell. This will result in poor
performance or damage.
NOTE
If an additional detector is added to the system, the
fluorescence detector should be the last detector in
the flow path except for evaporative detectors, like
LC-MSD. Otherwise, the back pressure generated by
the other detector may overload the flow cell and will
lead to a defective cell (maximum pressure is 20 bar
(2 MPa)).
Always use the outlet capillary set supplied with the
accessory kit.
8
Check for leaks: establish a flow and observe the flow
cell (outside of the cell compartment) and all capillary
connections.
9
Close the doors of the module.
Perform a Wavelength Calibration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals