Technical Information
MS120 Single Stage Rotary Vane Pumps User Manual / 87-900-150-01(B) 175/198
Section II
Connection to the Electric Supply
CAUTION!
It is recommended to connect the pump to the power supply through a dedicated
safety switch on the main electrical panel of the installation, or in proximity of
the power supply connection point.
WARNING!
The pump must be installed in a way that allows an easy interruption of the
line voltage.
Power supply cord: The correct cable for electrical wiring is a three
wires (Ph+N+Earth) cable. The wire section has to be at least 2 mm2
(AWG14).