EasyManuals Logo
Home>Agilent Technologies>Water Pump>MS40-S

Agilent Technologies MS40-S User Manual

Agilent Technologies MS40-S
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #182 background imageLoading...
Page #182 background image
Instructions for Use
182 MS40-S User Manual
Figure 19 MS40-S packaging
Installation
Do not install or use the pump in an environment exposed to atmospheric agents
(rain, snow, ice), dust, aggressive gases, or in explosive environments or those with
a high fire risk.
During operation, the following environmental conditions must be respected:
temperature: from +12 °C to +40 °C
relative humidity: 0 ÷ 95 % (non-condensing)
CAUTION
Before starting the pump, fill up with lubricating oil as the pump is delivered
empty.
CAUTION
Oil must be poured into the casing through the special threaded plughole and
NOT through the suction line.
NOTE
It is important that the oil level stays within the range MIN and MAX that is
indicated by the level gauge on the side of the pump.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MS40-S and is the answer not in the manual?

Agilent Technologies MS40-S Specifications

General IconGeneral
BrandAgilent Technologies
ModelMS40-S
CategoryWater Pump
LanguageEnglish

Related product manuals