MS40-S User Manual 201
Pump operations
Mains Connection
CAUTION
It is recommended to connect the pump to the power supply through a dedicated
Thermal magnetic circuit breaker (protection against short-circuit and line
overload).
Oil and Filter Cartridge Replacement Procedures
Oil Change Procedure
Remove the inlet and outlet pipes before opening the oil drain valve.
1 Place a tank for waste oil under the oil drain plug.
Figure 25 Oil Change
Oil drain valve
Fill oil cap

Table of Contents