Technical Information
MS40-S User Manual 215
Table 16 Window meanings
364
N
R
LAST WORKING HOUR AVERAGE ELECTRICAL POWER TO MOTOR [W]
365
N
R
LAST WORKING HOUR AVERAGE FREQUENCY [Hz]
382
N
R
AVERAGE TEMPERATURE [°C]
384
N
R
AVERAGE POWER DELIVERED TO THE MOTOR [W]
406
A
R
FIRMWARE PROGRAM LISTING CODE & REVISION
407
A
R
FIRMWARE PARAMETER LISTING CODE & REVISION
503
N
RW
RS485 SERIAL ADDRESS SETTING [0÷31] 0
504
L
RW
SERIAL COMMUNICATION SELECT (0=RS232; 1=RS485)
407
A
R
PARAMETER LISTING CODE & REVISION
503
N
RW
RS485 SERIAL ADDRESS SETTING [0-31]
504
L
RW
SERIAL TYPE SELECT (0=RS232; 1=RS485)
Error codes
If an error condition occurs, win206 assumes a value other than 0 and precisely:
128: pump rotor blocked
64: short circuit on the motor circuit
32: the mains voltage was out of specification during pump operation
16: motor overtempt
8: the mains voltage was too low (out of specification) when the pump was
started
4: controller over temperature
2: internal error (internal DC-Bus Undervoltage)
1: motor overcurrent
If more than one error occurs at the same time, the error code displayed will be the
sum of the codes of the individual errors.