EasyManuals Logo
Home>Agilent Technologies>Water Pump>MS40-S

Agilent Technologies MS40-S User Manual

Agilent Technologies MS40-S
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
Technical Information
MS40-S User Manual 215
Table 16 Window meanings
364
N
R
LAST WORKING HOUR AVERAGE ELECTRICAL POWER TO MOTOR [W]
365
N
R
LAST WORKING HOUR AVERAGE FREQUENCY [Hz]
382
N
R
AVERAGE TEMPERATURE [°C]
384
N
R
AVERAGE POWER DELIVERED TO THE MOTOR [W]
406
A
R
FIRMWARE PROGRAM LISTING CODE & REVISION
407
A
R
FIRMWARE PARAMETER LISTING CODE & REVISION
503
N
RW
RS485 SERIAL ADDRESS SETTING [0÷31] 0
504
L
RW
SERIAL COMMUNICATION SELECT (0=RS232; 1=RS485)
407
A
R
PARAMETER LISTING CODE & REVISION
503
N
RW
RS485 SERIAL ADDRESS SETTING [0-31]
504
L
RW
SERIAL TYPE SELECT (0=RS232; 1=RS485)
Error codes
If an error condition occurs, win206 assumes a value other than 0 and precisely:
128: pump rotor blocked
64: short circuit on the motor circuit
32: the mains voltage was out of specification during pump operation
16: motor overtempt
8: the mains voltage was too low (out of specification) when the pump was
started
4: controller over temperature
2: internal error (internal DC-Bus Undervoltage)
1: motor overcurrent
If more than one error occurs at the same time, the error code displayed will be the
sum of the codes of the individual errors.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MS40-S and is the answer not in the manual?

Agilent Technologies MS40-S Specifications

General IconGeneral
BrandAgilent Technologies
ModelMS40-S
CategoryWater Pump
LanguageEnglish

Related product manuals