Technical Information
208 MS40-S User Manual
Status LED
Table 12 Status LED
LED STATUS
CONTROLLER STATUS
Off
Stop
Green blinking
Ramp Autotuning
Green
Normal Operation
Red
Fail
Orange
Initial self-test
Electronic Self-Test
When the pump is switched on with the main On/Off switch the pump doesn’t start
immediately but it starts only after few seconds. During this phase the Status LED
remains lit orange fixed.
Rotor Lock Test
During the power operation, the rotational frequency is checked continuously. If the
rotor remains still for more than about 60 sec. the pump status is changed to “Fail”.
Rotor Blocked Error.