EasyManua.ls Logo

Agilent Technologies MS40-S

Agilent Technologies MS40-S
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use
186 MS40-S User Manual
NOTE
When the gases to be pumped out contain dust, it is advisable to insert a dust filter
before the inlet flange.
NOTE
When the gases to be pumped out contain large quantities of vapor, it is advisable
to include a condense separator before the inlet flange.
To make best use of the pump’s capacity, use only short, straight piping, with a
diameter not smaller than that of the inlet flange.
NOTE
If rigid piping is used, it is good practice to use a flexible joint in order to avoid
undue forcing of the connection on the pump.
The exhaust gases must be conveyed to the outside by connecting the exhaust port
to a dedicated pipe.
NOTE
An internal oil mist eliminator avoids pollution of the surrounding atmosphere by
the oil present in the exhaust duct during pump operation.
CAUTION
Never block the pump exhaust line. This would cause overpressure in the casing
with the risk of breaking the oil tank.

Table of Contents

Related product manuals