4-36 Service Guide N5242-90001
Troubleshooting PNA Series Microwave Network Analyzers
Measurement System Troubleshooting N5242A
reconnect the cable to the bridge and continue testing at “Checking the A34, A35, A36,
and A37 60-dB Source Step Attenuators” on page 4-36
7. If the 1 GHz signal is present and the analyzer does not have source attenuators and
bias tees, replace the test port coupler. Refer to “Removing and Replacing the A29–A32
Test Port Couplers” on page 7-38.
Checking the A34, A35, A36, and A37 60-dB Source Step Attenuators
1. Locate the appropriate semirigid cable at the output of the source step attenuator to be
checked:
Options 219, 224, 419, and 423
Port 1; W72 of A34
Port 2; W84 of A37
Port 3; W76 of A35
Port 4; W80 of A36
2. Using a 5/16-inch torque wrench, disconnect the semirigid cable at the step attenuator.
3. Connect the spectrum analyzer to the open step attenuator connector. Set the spectrum
analyzer to measure a signal at 1 GHz.
4. If the 1 GHz signal is not present, replace the source step attenuator. Refer to
“Removing and Replacing the A34–A37 Source Attenuators, the A38–A41 Bias Tees,
and the A42–A45 Receiver Attenuators” on page 7-42.
5. If the 1 GHz signal is present, replace the associated bias tee. Refer to “Removing and
Replacing the A34–A37 Source Attenuators, the A38–A41 Bias Tees, and the A42–A45
Receiver Attenuators” on page 7-42.
Checking the A46, A47, A48, and A49 Source Bypass Switches
1. Locate the appropriate semirigid cable at the output of the source bypass switch to be
checked:
Options 224 and 423
Port 1; W96 of A46
Port 2; W110 of A49
Port 3; W102 of A47
Port 4; W106 of A48
2. Using a 5/16-inch torque wrench, disconnect the semirigid cable at the bypass switch.
3. Connect the spectrum analyzer to the open bypass switch connector. Set the spectrum
analyzer to measure a signal at 1 GHz.
4. If the 1 GHz signal is not present, replace the source bypass switch. Refer to “Removing
and Replacing the A46–A49 Bypass Switches and the A50 Combiner” on page 7-46.
5. If the 1 GHz signal is present, replace the associated bridge. Refer to “Removing and
Replacing the A25–A28 Bridges and Bridge Mounting Brackets” on page 7-36.