7-50 Service Guide N5242-90001
Repair and Replacement Procedures PNA Series Microwave Network Analyzers
Removing and Replacing the A51 Hard Disk Drive (HDD) N5242A
Reinstalling the HDDA
1. Reinstall the A51 hard disk drive into the analyzer:
a. Connect the ribbon cable (item
) to the new A51 hard disk drive assembly.
b. Place the CPU board assembly top cover on the A51 hard disk drive assembly and
install the four mounting screws (item
).
c. Position the top cover (with the A51 hard disk drive assembly attached) over the A17
CPU board assembly and reconnect the ribbon cable (item
) to the CPU board.
d. Align the top cover and install the fifteen attachment screws (item
).
Set Up the Agilent Administrator Password
If the hard disk has been replaced, it is necessary to set up the Agilent administrator
password for the analyzer. This requires obtaining a password from Agilent:
Obtaining a Password
If this procedure is being performed by Agilent personnel, go to the following Internet
site and follow the instructions under “Agilent Administrator Password”.
http://mktwww.soco.agilent.com/field/service/network/pna/support.htm
If this procedure is being performed by non-Agilent personnel, contact Agilent for a
password before beginning this procedure. Refer to “Contacting Agilent” on page 2-13.
Entering Your Agilent Administrator Password
The network analyzer must be powered up and operating for this procedure.
1. On the Windows Desktop, right-click on My Computer, and then click Manage.
2. In the Computer Management window, navigate to System Tools, Local Users and
Groups, Users.
3. Right-click on Agilent, and then click on Set Password.
a. Type the password obtained from Agilent as the Password. Refer to “Obtaining a
Password” in this procedure for information if you have not yet obtained this
password.
b. Type the password again in the Confirm password block.
c. Click OK.
4. Right-click on Agilent, and then click on Properties.
a. Unselect User must change password at next logon (if not already unselected).
b. Unselect User cannot change password (if not already unselected).
c. Select Password never expires (if not already selected).
d. Click OK.
5. Close the Computer Management window.