Service Guide N5242-90001 3-11
PNA Series Microwave Network Analyzers Tests and Adjustments
N5242A Preliminary Checks
5. The result of the operators check will be shown as a PASS or FAIL next to each test
(refer to Figure 3-3). The PNA Operators Check dialog box will look different for
different PNA model numbers and installed options. Some of the tests are performed
only if the appropriate options are installed in the PNA.
Figure 3-3 Operators Check Dialog Box
If the Operators Check Fails
1. Clean the test ports, shorts, and adapters. Torque to specification. Repeat the check.
2. If the check still fails, suspect a faulty component. Refer to “Measurement System
Troubleshooting” on page 4-23 to begin troubleshooting to determine the faulty
component.