Appendix A Specifications
126 Series N6700 User’s Guide
Agilent N6700A/B MPS Mainframe
N6700A/B
Command Processing Time:
From receipt of command to start of the
output change
1 ms
Protection Response Characteristics:
INH input
Fault on coupled outputs
5 µs from receipt of inhibit to start of shutdown
< 10 µs from receipt of fault to start of shutdown
Digital Control Characteristics:
NOTE 1
Maximum voltage ratings
Pins 1 and 2 as FLT output
Pins 1 - 7 as digital/trigger outputs
(pin 8 = common)
Pins 1 - 7 as digital/trigger inputs and
pin 3 as INH input
(pin 8 = common)
+16.5 VDC/ 5 VDC between pins (pin 8 is internally connected to chassis ground).
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 0.14 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA; 1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Maximum low-level input voltage = 0.8 V
Minimum high-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Interface Capabilities:
GPIB:
USB 2.0:
10/100 LAN:
Built-in Web server
SCPI - 1993, IEEE 488.2 compliant interface
Requires Agilent IO Library version M.01.01 and up, or 14.0 and up
Requires Agilent IO Library version L.01.01 and up, or 14.0 and up
Requires Internet Explorer 5+ or Netscape 6.2+
Environmental Conditions
Operating environment
Temperature range
Relative humidity
Altitude
Storage temperature
LED statement
Indoor use, installation category II (for AC input), pollution degree 2
0°C to 55°C (current is derated 1% per °C above 40°C ambient temperature)
Up to 95%
Up to 2000 meters
-30°C to 70°C
Any LEDs used in this product are Class 1 LEDs as per IEC 825-1
1
Agilent N6700A mainframes have only 4 digital control pins. Refer to Appendix D for pin assignments.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N6751A and is the answer not in the manual?

Agilent Technologies N6751A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN6751A
CategoryPower Supply
LanguageEnglish

Related product manuals