Service Appendix D
Series N6700 User’s Guide 187
Device-dependent Errors (continued)
200 Hardware error channel <channel>
A hardware error has occurred on the specified channel.
201 Invalid configuration, empty slots
There is an empty slot between modules. This configuration is not allowed.
202 Selftest Fail
A selftest failure has occurred. See selftest failure list for details.
203 Compatibility function not implemented
The requested compatibility function is not available.
204 NVRAM checksum error
A checksum error has occurred in the instrument’s nonvolatile random access memory.
205 NVRAM full
The nonvolatile random access memory of the instrument is full.
206 File not found
The internal calibration file or the internal channel attribute file was not found in NVRAM.
207 Cal file version error
The calibration file was written or read using old firmware. Firmware must be updated.
302 Option not installed
The option that is programmed by this command is not installed.
303 There is not a valid acquisition to fetch from
There is no valid data in the measurement buffer.
304 Volt and curr in incompatible transient modes
Voltage and current cannot be in Step and List mode at the same time.
305 A triggered value is on a different range
A triggered value is on a different range than the one that is presently set.
306 Too many list points
Too many list points have been specified.
307 List lengths are not equivalent
One or more lists are not the same length.
308 This setting cannot be changed while transient trigger is initiated
Setting cannot be changed while the instrument is waiting for or executing a trigger sequence.
309 Cannot initiate, voltage and current in fixed mode
Cannot initiate the transient generator because either the voltage or the current function is set
to Fixed mode.
Command Errors (these errors set Standard Event Status register bit #5)
100 Command error
Generic syntax error.
101 Invalid character
An invalid character was found in the command string.
102 Syntax error
Invalid syntax was found in the command string. Check for blank spaces.
103 Invalid separator
An invalid separator was found in the command string. Check for proper use of , ; :
104 Data type error
A different data type than the one allowed was found in the command string.
105 GET not allowed
A group execute trigger is not allowed in a command string.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N6751A and is the answer not in the manual?

Agilent Technologies N6751A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN6751A
CategoryPower Supply
LanguageEnglish

Related product manuals