Appendix E Compatibility
194 Series N6700 User’s Guide
Agilent 662xA Command Description Similar SCPI
Command
IHI <ch>
Generates error 113. Unit must be calibrated using SCPI commands.
see Appendix B
ILO <ch>
Generates error 113. Unit must be calibrated using SCPI commands.
see Appendix B
IOUT? <ch> Queries the measured output current of the specified output channel. The
response is a real number. The front panel display can be used to monitor
the measured output current (and voltage) of the selected output channel.
MEAS:CURR?
IRLO <ch>
Generates error 113.
None
IRHI <ch>
Generates error 113.
None
IRLN <ch>
Generates error 113.
None
IRHN <ch>
Generates error 113.
None
IRSET <ch>, <x> Sets the full-scale current range programming for an output channel. The
power system will select the lowest operating range that will contain the
value (x). If the present operating current is greater than the full-scale
range selected, the present output current will be reduced to the full-scale
value of the new range selected. If the current entered is larger than the
highest possible range, an error will result.
SENS:CURR:RANG
IRSET? <ch> Queries the full-scale current range setting of an output channel.
SENS:CURR:RANG?
ISET <ch>, <current> Sets the current of the specified output channel.
CURR
ISET? <ch> Queries the present current setting of the specified output channel. The
response is a real number.
CURR?
ISTEP <ch>, <x> Increases or decreases the output channel current in steps. If the step
value < x > is positive, output current will increase by the step value, if the
step value is negative, the output current will decrease by the step value.
Stepping beyond the active operating range causes an error.
LIST:CURR
METER <ch> Does nothing.
None
METER? Always returns a 1.
None
OCP <ch>, <on/off> Enables the over-current protection circuit for the specified output channel.
This circuit, when enabled, causes the output to go to the off state when
the output is in + CC mode. On/off is a 1 to turn on (enable) or a 0 to turn
off (disable) the circuit.
CURR:PROT:STAT
OCP? < ch > Queries the over-current protection circuit on/off status for the specified
output channel. Response is either a 1 (on) or a 0 (off).
CURR:PROT:STAT?
OCRST < ch > Returns the specified output channel to its previous settings after it has
been turned off by a protection shutdown.
OUTP:PROT:CLE
OUT < ch >, < on/off > Turns the specified output channel on or off. <on/off> equals 1 to turn the
output on and equals 0 to turn the output off.
OUTP
OUT? < ch > Queries whether the specified output channel is turned on or off . The
response is either 1 (on) or O (off).
OUTP?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N6751A and is the answer not in the manual?

Agilent Technologies N6751A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN6751A
CategoryPower Supply
LanguageEnglish

Related product manuals