EasyManuals Logo

Agilent Technologies U8002A User Manual

Agilent Technologies U8002A
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
Service Guide 5
U8001A/U8002A User’s and Service Guide 53
Enable Auto- triggering
8 Allow the scope run for a few seconds to generate enough
measurement points.
9 Obtain the maximum peak- to- peak voltage measurement as
indicated in the scope. Divide this value by 10 to get the CV
peak- to- peak noise measurement. The result should not
exceed 12 mV
pp
.
10 Configure the RMS voltmeter as below:
Set high pass filter to10 Hz
AC coupling
11 Disconnect the oscilloscope and connect an RMS voltmeter in
its place. Do not disconnect the 50
Ω termination. Divide the
reading of the RMS voltmeter by 10. The result should not
exceed the RMS limits of 1 mV
RMS
.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U8002A and is the answer not in the manual?

Agilent Technologies U8002A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU8002A
CategoryPower Supply
LanguageEnglish

Related product manuals