26 U8001A/U8002A User’s and Service Guide
3 Operations and Features
Programming Over Current Protection
Over current protection guards the load against output currents
reaching values greater than the programmed protection level.
The following steps show how to enable and disable the over
current protection (OCP), how to set the OCP trip level and how
to clear the over voltage condition.
To set the OCP trip level and enable OCP
1 Press on the “Over Current” button.
• The LIMIT annunciator will be turned off.
• The OCP annunciator will fast- blink continuously.
• The display will show the OCP value and the character “A”
will fast- blink continuously.
2 Turn the knob to adjust the OCP value.
3 Press on the “Over Current” button again to end the
adjustment and enable the OCP.
• Display will show “donE”.
• The OCP annunciator will be turned on.