EasyManuals Logo

Agilent Technologies U8002A User Manual

Agilent Technologies U8002A
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
54 U8001A/U8002A User’s and Service Guide
5Service Guide
Load Transient Response Time
This test measures the time for the output voltage to recover to
within 15 mV of nominal output voltage following a load change
from full load to half load or vice versa.
Procedures:
1 Power off the power supply and connect the output to be
tested as shown in Figure 5- 2(A) with an oscilloscope.
Operate the electronic load in constant current mode.
2 Power on the power supply.
3 When the display is in the limit mode, program the output
voltage to full rated value, i.e. 30 V and output current to the
maximum programmable value.
4 Enable the output.
5 Set the electronic load to transient operation mode between
one half of the output's full rated value and the output’s full
rated value at a 1 kHz rate with 50% duty cycle.
6 Set the oscilloscope coupling, internal sync and lock on
either the positive or negative load transient.
7 Adjust the oscilloscope to display transients as shown in
Figure 5- 4. Note that the pulse width (t2–t1) of the
transients at 15 mV from the base line is no more than 50 µs
for the output.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U8002A and is the answer not in the manual?

Agilent Technologies U8002A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU8002A
CategoryPower Supply
LanguageEnglish

Related product manuals