10. Maintenance
The displaying unit shall be subject to preventive inspection prior to use according to section
7.1. and shall be cleaned and disinfected according to chapter 9.
No other preventive inspection, maintenance or calibration activities are required for the
displaying unit.
11. Disposal
At the end of product life, clean and disinfect the displaying unit (see chapter 9).
Before disposing of the displaying unit, it is advised to follow these steps to erase all data:
• Log in as Administrator.
• Delete all files stored in the displaying unit from the Archive
- Select all folders and
press the bin
.
• Delete Wi-Fi network and password though the Settings
– Setup – Network setup
menu.
• Delete DICOM network via the Settings
– Setup – DICOM setup menu.
• Delete all Advanced Users from the Settings
– User profiles menu.
Then, dispose of the displaying unit in accordance with local guidelines for electrical and
electronic waste.
12. Technical Product Specifications
12.1. Standards Applied
The displaying unit function conforms with:
• IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety
and essential performance.
• IEC 60601-1-2 Medical electrical equipment – Part 1-2 General requirements for safety –
Collateral standard: Electromagnetic compatibility - Requirements for test.
• IEC 60601-2-18 Medical electrical equipment – Part 2-18: Particular requirements for
the basic safety and essential performance of endoscopic equipment.
12.2. Specifications for the Displaying Unit
Display
Resolution 1920 x 1080 pixels
Orientation Landscape
Display type 12.8” colour TFT LCD
Startup time Live image available within 8 seconds
Connections
USB connections Front: 1 x USB 3.0 type A
Back: 3 x USB 3.0 type A and 2 x USB 2.0 type A
Video output DVI-D (1920 x 1080p, 60 fps) and 3G-SDI (1920 x
1080p, 60 fps)
Wi-Fi Supports IEEE Standards 802.11ac/a/b/g/n
LAN Ethernet RJ45 connector, 10/100/1000 Mbps
Trigger output 2 x D-SUB9 and 2 x 3.5 mm Jack
Connection to potential equalization
cable
DIN 42801 standard
Memory
42