Guidance and manufacturer’s declaration – electromagnetic emission
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment Guidance
RF emission
CISPR 11
Group 1 The system uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class A The system is suitable for use in all establishments
other than domestic, and may be used in domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes, ,
provided the following NOTE 1 is heeded.
Harmonic
emission IEC/EN
61000-3-2
Complies
Voltage
fluctuations /
flicker emissions
IEC/EN 61000-3-3
Complies
NOTE 1: The emissions Characteristics of this equipment makes it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR
11 class B is normally required) this equipment might not offer adequate protection to the
radio-frequency communication service. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
Guidance and manufacturer’s declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Immunity test IEC 60601-1-2
test level
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
+/- 8 kV
contact
+/- 2, 4, 8, 15
kV air
If floors are covered with
synthetic material the
relative humidity shall be
least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for input /
output lines
+/- 2 kV
power supply
lines N/A
Mains power quality shall
be that of a typical
commercial or hospital
environment.
48