Modbus Communication Interface | A-9
TABLE A-1 cont.
REG DEFINITION UNITS SCALING ACCESS
85 Read only except test mode (AnalogInput= 4E Hex) - 10 r,w
86 Read only except test mode (AnalogOutput= 4F Hex - 1 r,w
87 Used in test mode (ReadTestFixture= 50 Hex)
-
1 r
88 Sample = 0,generator = 1,reference = 2
-
1 r,w
(ValveState= 56 Hex)
89 Current hour (Hour= 57 Hex) hour 1 r,w
90 Current minute (Minute= 58 Hex) min 1 r,w
91 Current second (Second= 59 Hex) sec 1 r,w
92 Current month (Month= 5A Hex) - 1 r,w
93 Current day (Day= 5B Hex) - 1 r,w
94 Current year (Year= 5C Hex) - 1 r,w
95 Used in test only (WriteTestFixture= 63 Hex) - 1 -
96 Current day of week (WeekDay= 6D Hex) - 1 r,w
97 Verification coefficient (MoistureSpan= 6E Hex) - 1000 r,w
98 Track or hold current output flag (HoldOut= 71 Hex) - 1 r,w
99 Fast cycle =0, slow cycle=1 (SlowTiming= 72 Hex) - 1 r,w
100 Minimum verification duration min 1 r,w
(VerDurationLimit= 73 Hex)
101 Adjust span in verification flag (AdjustSpan= 74 Hex) - 1 r,w
102 Alarm 1 enable flag (EnableAlarm1= 75 Hex) - 1 r,w
103 Alarm 2 enable flag (EnableAlarm2= 76 Hex) - 1 r,w
104 Reserved - - -
105 Dryer limit (DryerPpmHours= 7A Hex) ppmH 1 r,w
106 Dryer counter (CurrentPpmHours= 7B Hex) - 1 r,w
107 Flow correction coefficient (FlowCorrection= 8B Hex) - 10 r,w
108 Corrected generator reading (MoistureCorr= 8C Hex) - 10 r,w
109 Flow correction multiplier (FlowWeight= 8D Hex) - 10 r,w
110 Previous span before adjustment - 1000 r,w
(OldMoistSpan= 8F Hex)
111 Skip span drift limit (IgnoreSpanDrift= 90 Hex) - 1 r,w
112 Previous offset number (OldFreqOffset= 91 Hex) - 1000 r,w