6-6 | Model 5100 HD Analyzer
29
116
r
int16
MANCMD
82
1
Parameter used to request a state
machine change. 1=Not ready,
2=Measure, 3=Zero, 4=Span, 5=Zero
Hold, 6=Span Hold, 7=Idle
117
rw
char[20]
CONCUNIT
83
10
Concentration Unit of Measure for
analyte 1
118
rw
int16
FLOATFRMT
93
1
Float Format index. 0=comma as
decimal, 1=period as decimal
119
rw
uint16
IDATEFRMT
94
1
International date format index.
0=MM\DD\YYYY,1=MM-DD-YYYY,
2=DD\MM\YYYY,3=DD-MM-YYYY
120
rw
char[10]
PRESUNIT
95
5
Pressure Unit of Measure, case
sensitive. Valid strings include psi,
psia, inH20, inHg, bar, cmH20,
kg/cm2, kPa, mbar, micron, mmHg,
oz/in2, torr, atm
121
rw
char[10]
TEMPUNIT
100
5
Temp Unit of Measure, case
sensistive. Valid strings include
degreeC, degreeF, degreeK
122
r
uint
DATAVALID
105
2
Data Valid. 0=Not Valid, 1=Valid
123
r
int16
DMODE
107
1
0=Line-lock mode, 1=Sample,
2=Calib
124
rwf
int16
COMPPT
108
1
Pressure and temp compensation.
0=none, 1=pressure, 2=temperature,
3=both
125
r
uint
SYSALARM1
109
2
Bit-mapped system alarms
126
r
uint
SYSALARM2
111
2
Bit-mapped system alarms
127
r
uint
USERALARM
113
2
Bit-mapped system alarms
128
r
uint
CONCALARM
115
2
Bit mapped concentration alarms
129
r
int16
SYSSTATUS
117
1
SystemStatus.0=Everythingok,Not
0=something is wrong
130
r
int16
SYSSTATE
118
1
SystemState.100=NotReady,
201=MeasureRun,502=Idle,
600=diag
131
r
uint
TOTUPTIME
119
2
Total uptime in seconds
132
r
uint
UPTIME
121
2
Uptime in seconds
133
r
float
CELLTEMPD
123
2
Cell temp for display
134
r
float
ANAOUT1
125
2
CurrentOutput1
135
rw
float
ANAOUT1O
127
2
CurrentOutput1Offset
136
rw
uint16
ANAOUT1P
129
1
CurrentOutput1Parameter
137
rw
float
ANAOUT1S
130
2
CurrentOutput1Span
138
rw
int16
ANAOUT1T
132
1
CurrentOutput1OutputType.0=4-
20 mode, 1=0-20 mode.
Ifoverangeenable(ANAOUT1OR=
1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or
low values can also be set by fault
value,SeeFaultoption(ANAOUT1F)
139
rw
uint16
ANAOUT1OR
133
1
CurrentOutput1OverrangeEnable.
0=disabled, 1=enabled
140
rw
uint16
ANAOUT1F
134
1
CurrentOutput1FaultOption.0=No
fault,1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT1T=0):
Lowendofrangebecomes3mA
Outputratiobecomes 0.
30
141
r
float
ANAOUT2
135
2
CurrentOutput2-
142
rw
float
ANAOUT2O
137
2
CurrentOutput2Offset-
143
rw
uint16
ANAOUT2P
139
1
CurrentOutput2Parameter-
144
rw
float
ANAOUT2S
140
2
CurrentOutput2Span-
145
rw
int16
ANAOUT2T
142
1
CurrentOutput2OutputType.0=4-
20 mode, 1=0-20 mode.
Ifoverangeenable(ANAOUT2OR=
1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or
low values can also be set by fault
value,SeeFaultoption(ANAOUT2F)
-
146
rw
uint16
ANAOUT2OR
143
1
CurrentOutput2 OverrangeEnable.
0=disabled, 1=enabled
147
rw
uint16
ANAOUT2F
144
1
CurrentOutput2FaultOption.0=No
fault,1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT2T=0):
Lowendofrangebecomes3mA
Outputratiobecomes0.
148
r
uint16
RELAY1
145
1
Relay1.0=OPEN,1=CLOSED
149
rw
uint16
RELAY1D
146
1
Relay1Delay.Holdstherelayoutput
for this set amount of time.
150
rw
uint16
RELAY1F
147
1
Relay1Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
151
rw
uint16
RELAY1S
148
1
Relay1Source.Thesource
parameter number used to evaluated
the relay function.
152
rw
float
RELAY1V
149
2
Relay1Value.Thevalueusedto
evaluate the relay function.
153
r
uint16
RELAY2
151
1
Relay2.0=OPEN,1=CLOSED
154
rw
uint16
RELAY2D
152
1
Relay2Delay.Holdstherelayoutput
for this set amount of time.
155
rw
uint16
RELAY2F
153
1
Relay2Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
156
rw
uint16
RELAY2S
154
1
Relay2Source.Thesource
parameter number used to evaluated
the relay function.
157
rw
float
RELAY2V
155
2
Relay2Value.Thevalueusedto
evaluate the relay function.
158
r
uint16
RELAY3
157
1
Relay3.0=OPEN,1=CLOSED
159
rw
uint16
RELAY3D
158
1
Relay3Delay.Holdstherelayoutput
for this set amount of time.
160
rw
uint16
RELAY3F
159
1
Relay3Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
161
rw
uint16
RELAY3S
160
1
Relay3Source.Thesource
parameter number used to evaluated
the relay function.
162
rw
float
RELAY3V
161
2
Relay3Value.Thevalueusedto
evaluate the relay function.
163
r
uint16
RELAY4
163
1
Relay4.0=OPEN,1=CLOSED