6-6 | Model 5100 HD Analyzer
Parameter used to request a state
machine change. 1=Not ready,
2=Measure, 3=Zero, 4=Span, 5=Zero
Hold, 6=Span Hold, 7=Idle
Concentration Unit of Measure for
analyte 1
Float Format index. 0=comma as
decimal, 1=period as decimal
International date format index.
0=MM\DD\YYYY,1=MM-DD-YYYY,
2=DD\MM\YYYY,3=DD-MM-YYYY
Pressure Unit of Measure, case
sensitive. Valid strings include psi,
psia, inH20, inHg, bar, cmH20,
kg/cm2, kPa, mbar, micron, mmHg,
oz/in2, torr, atm
Temp Unit of Measure, case
sensistive. Valid strings include
degreeC, degreeF, degreeK
Data Valid. 0=Not Valid, 1=Valid
0=Line-lock mode, 1=Sample,
2=Calib
Pressure and temp compensation.
0=none, 1=pressure, 2=temperature,
Bit mapped concentration alarms
SystemStatus.0=Everythingok,Not
0=something is wrong
SystemState.100=NotReady,
201=MeasureRun,502=Idle,
CurrentOutput1Parameter
CurrentOutput1OutputType.0=4-
20 mode, 1=0-20 mode.
Ifoverangeenable(ANAOUT1OR=
1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or
low values can also be set by fault
value,SeeFaultoption(ANAOUT1F)
CurrentOutput1OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput1FaultOption.0=No
fault,1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT1T=0):
Lowendofrangebecomes3mA
Outputratiobecomes 0.
CurrentOutput2Offset-
CurrentOutput2Parameter-
CurrentOutput2OutputType.0=4-
20 mode, 1=0-20 mode.
Ifoverangeenable(ANAOUT2OR=
1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or
low values can also be set by fault
value,SeeFaultoption(ANAOUT2F)
CurrentOutput2 OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput2FaultOption.0=No
fault,1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT2T=0):
Lowendofrangebecomes3mA
Outputratiobecomes0.
Relay1.0=OPEN,1=CLOSED
Relay1Delay.Holdstherelayoutput
for this set amount of time.
Relay1Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay1Source.Thesource
parameter number used to evaluated
the relay function.
Relay1Value.Thevalueusedto
evaluate the relay function.
Relay2.0=OPEN,1=CLOSED
Relay2Delay.Holdstherelayoutput
for this set amount of time.
Relay2Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay2Source.Thesource
parameter number used to evaluated
the relay function.
Relay2Value.Thevalueusedto
evaluate the relay function.
Relay3.0=OPEN,1=CLOSED
Relay3Delay.Holdstherelayoutput
for this set amount of time.
Relay3Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay3Source.Thesource
parameter number used to evaluated
the relay function.
Relay3Value.Thevalueusedto
evaluate the relay function.
Relay4.0=OPEN,1=CLOSED