Maintenance | 6-7
CurrentOutput2Offset-
CurrentOutput2Parameter-
CurrentOutput2OutputType.0=4-
20 mode, 1=0-20 mode.
Ifoverangeenable(ANAOUT2OR=
1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or
low values can also be set by fault
value,SeeFaultoption(ANAOUT2F)
-
CurrentOutput2 OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput2FaultOption.0=No
fault,1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT2T=0):
Lowendofrangebecomes3mA
Outputratiobecomes0.
Relay1.0=OPEN,1=CLOSED
Relay1Delay.Holdstherelayoutput
for this set amount of time.
Relay1Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay1Source.Thesource
parameter number used to evaluated
the relay function.
Relay1Value.Thevalueusedto
evaluate the relay function.
Relay2.0=OPEN,1=CLOSED
Relay2Delay.Holdstherelayoutput
for this set amount of time.
Relay2Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay2Source.Thesource
parameter number used to evaluated
the relay function.
Relay2Value.Thevalueusedto
evaluate the relay function.
Relay3.0=OPEN,1=CLOSED
Relay3Delay.Holdstherelayoutput
for this set amount of time.
Relay3Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay3Source.Thesource
parameter number used to evaluated
the relay function.
Relay3Value.Thevalueusedto
evaluate the relay function.
Relay4.0=OPEN,1=CLOSED
Relay4Delay.Holdstherelayoutput
for this set amount of time.
Relay4Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
Relay4Source.Thesource
parameter number used to evaluated
the relay function.
Relay4Value.Thevalueusedto
evaluate the relay function.
Master enable for all relays.
0=Disabled,1=Enabled
Analyte1bandhighlimiteband1
Analyte1bandlowlimiteband1
Analyte1band high limite band 2
Analyte1bandlowlimiteband2
Analyte1bandhighlimiteband3
Analyte1bandlowlimiteband3
Analyte1bandhighlimiteband4
Analyte1bandlowlimiteband4
HeaterActuatorType.0=Reverse,
1=Forward.
Heater Deviation Tolerance. Ignored if
0.Otherwisehighorlowlimitalarm
are set when value is out bounds of 1
deviation
Cell Deviation Tolerance. Ignored if 0.
Otherwisehighorlowlimitalarmare
set when value is out bounds of 1
deviation
Calibration temperature for
concentration compensation.
Calibration pressure for concentration
compensation
Interval in minutes to save the log
data
Concentration Unit of Measure for
analyte 2
Decimal places to use for displaying
analyte 1