6-8 | Model 5100 HD Analyzer
31
164
rw
uint16
RELAY4D
164
1
Relay4Delay.Holdstherelayoutput
for this set amount of time.
165
rw
uint16
RELAY4F
165
1
Relay4Function.0=Disabled,
1=Concentration, 3=Data Valid,
4=System alarm
166
rw
uint16
RELAY4S
166
1
Relay4Source.Thesource
parameter number used to evaluated
the relay function.
167
rw
float
RELAY4V
167
2
Relay4Value.Thevalueusedto
evaluate the relay function.
168
rwf
boolean
RELAYENA
169
1
Master enable for all relays.
0=Disabled,1=Enabled
169
rwf
char[20]
ALYTE1
170
10
Analyte1Name
170
r
float
ALYTE1CON
180
2
Analyte1Concentration
171
rw
float
CONCLIMITH1
182
2
Analyte1bandhighlimiteband1
172
rw
float
CONCLIMITL1
184
2
Analyte1bandlowlimiteband1
173
rw
float
CONCLIMITH2
186
2
Analyte1band high limite band 2
174
rw
float
CONCLIMITL2
188
2
Analyte1bandlowlimiteband2
175
rw
float
CONCLIMITH3
190
2
Analyte1bandhighlimiteband3
176
rw
float
CONCLIMITL3
192
2
Analyte1bandlowlimiteband3
177
rw
float
CONCLIMITH4
194
2
Analyte1bandhighlimiteband4
178
rw
float
CONCLIMITL4
196
2
Analyte1bandlowlimiteband4
179
rwf
char[20]
ALYTE2
198
10
Analyte2Name
180
r
float
ALYTE2CON
208
2
Analyte2Concentration
181
r
float
TECTEMP
210
2
TECTemperatureValue
182
rwf
int16
TECAT
212
1
HeaterActuatorType.0=Reverse,
1=Forward.
183
rwf
float
TECDEVT
213
2
Heater Deviation Tolerance. Ignored if
0.Otherwisehighorlowlimitalarm
are set when value is out bounds of 1
deviation
184
rwf
float
TECDT
215
2
Heater Deriviative Time
185
r
float
TECHDC
217
2
Heater Duty Cycle
186
rwf
float
TECIT
219
2
Heater Integral Time
187
rwf
float
TECPB
221
2
Heater Proportional Band
188
rwf
float
TECSP
223
2
Heater Set Point
189
rwf
float
TECST
225
2
TECSampleTime
190
rwf
float
TECHYS
227
2
TEChysterisi
191
r
float
ELECTEMP
229
2
ElectronicsTemperature
192
r
float
CELLTEMP
231
2
CellRTD1Filteredvalue
193
rwf
float
CELLSP
233
2
Cell Set Point
194
rwf
float
CELLDEVT
235
2
Cell Deviation Tolerance. Ignored if 0.
Otherwisehighorlowlimitalarmare
set when value is out bounds of 1
deviation
195
rwf
float
NORMTEMP
237
2
Calibration temperature for
concentration compensation.
196
rwf
float
NORMPRES
239
2
Calibration pressure for concentration
compensation
197
rwf
float
MAXTEMP
241
2
MaxTECtemperature
198
rwf
float
MINTEMP
243
2
MinTECtemperature
199
rw
uint
LOGSAVEINT
245
2
Interval in minutes to save the log
data
200
rw
char[10]
CONCUNIT2
247
5
Concentration Unit of Measure for
analyte 2
201
rw
uint
ALYTE1RES
252
2
Decimal places to use for displaying
analyte 1
32
202
rw
uint
ALYTE2RES
254
2
Decimal places to use for displaying
analyte 2
203
rw
boolean
RELAY1NC
256
1
Normally closed indicator for relay 1.
1=Normally closed, 0=Normally open
204
rw
boolean
RELAY2NC
257
1
Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
205
rw
boolean
RELAY3NC
258
1
Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
206
rw
boolean
RELAY4NC
259
1
Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
207
rw
uint16
MENUTO
260
1
Menu timeout. The time that the menu
will navigate to the homescreen after
no activity. 0 is disabled
208
rw
uint16
PWDTO
261
1
Password timeout. The time that a
user logged into the front display
menu will be logged out after no
activity. 0 is disabled
209
r
uint
SYSWARN1
262
2
Temporary cache to miror warnings in
SYSALARM1
210
r
uint
SYSWARN2
264
2
Temporary cache to miror warnings in
SYSALARM2
211
r
uint
SYSSTWARN
266
2
system statuc warning
212
rw
uint16
PTSEED
268
1
TheseedvalueforallIIRfilters.
213
rwf
float
MAXTEMP_2
269
2
MaxTECtemperature2
214
rwf
float
MINTEMP_2
271
2
MinTECtemperature2
215
rw
float
REFCONC_2
273
2
ConcentrationofReferencecall
216
rwf
int16
STATICPT
275
1
Pressure(STCELLP)andtemp
(STCELLT)staticvalueusage.
0=neither, 1=use static pressure,
2=use static temperature, 3= use both
217
rwf
float
STCELLT
276
2
Static cell temperature. Use by
settingSTATICPTto2or3
218
rwf
float
STCELLP
278
2
Static cell pressure. Use by setting
STATICPTto1or3
219
rw
int16
ANAOUT1H
280
1
Analogoutput1,0- follow source
during cal, 1 - hold output during cal
220
rw
int16
ANAOUT2H
281
1
Analogoutput2,0- follow source
during cal, 1 - hold output during cal
L82
221
rwf
float
LASTH_2
282
2
Threshold in 2f signal
222
rwf
float
LASREFPWR_2
284
2
ReferenceLaserPower(powerat
calib)
223
rwf
float
LASTS_2
286
2
LaserSamplePeriodinseconds
224
rwf
float
LASPWR_2
288
2
LaserPower
225
rwf
float
CELLLENGTH
290
2
Cell length, concentrations are
multiplied by this parameter
226
r
int16
DIAGS
292
1
Used to change state to diagnostics
from the menu
227
rw
float
CONCMIN_2
293
2
Minimum Concentration
228
rw
float
CONCMAX_2
295
2
Maximum Concentration
229
rwf
int16
PASSCODE
297
1
PASSCODEtochangesystem
params
230
rwf
float
LASTH
298
2
Threshold in 2f signal
231
rwf
float
LASREFPWR
300
2
ReferenceLaserPower(powerat
calib)
232
rwf
float
LASTS
302
2
LaserSamplePeriodinseconds
233
rwf
float
LASPWR
304
2
LaserPower