EasyManua.ls Logo

Ametek 931S

Ametek 931S
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-4 | Model 931S / Model 932S UV Analyzers
Frequency Component/Task
Every Year Measuring Cell
Replace the o-rings and Cell Windows and clean the interior
surfaces to remove any contaminants.
This schedule is a minimum requirement. If other conditions are
found to be present, as discussed in “Measuring Cell Maintenance,”
more frequent cleaning may be required.
Every 2 Years Chopper Assembly, Optical Bench – Model 932S Analyzers
Replace the Chopper Motor Drive Belt (o-ring) and Chopper Wheel
Bearings. If there are recurring analyzer alarms that suggest
problems with the Chopper Wheel or other related parts in the
Chopper Assembly, replace these parts sooner (alarms: “w: PMT
signal out of range, f: Filter wheel speed out of range,
w: Excessive zero drift”, w: ALC out of range, w: Bench
auto-setup problem).
See “Chopper Assembly Maintenance – Model 932S Analyzers
Only” in this chapter.
Other Flamepath Gap
During each analyzer maintenance, use a feeler gauge to check the
amepath gap of the following locations. The gap must not exceed
the listed gap for each location; if the gap exceeds this value,
contact AMETEK for advice. See “Examining and Caring For the
Flamepaths” in this chapter.
Upper/Lower Enclosure anges (enclosure door and housing
joining surfaces).
Maximum Flamepath Gap: 0.1 mm (Group IIB)
0.04 mm (Group IIB+H2)
Cover, Communications Port (Communications Port Body-Cover
joining surfaces)
Maximum Flamepath Gap: 0.1 mm (Group IIB)
0.04 mm (Group IIB+H2)
Cell Heater Block and Cell Heater Block Lid joining surfaces.
Maximum Flamepath Gap: 0.1 mm (Group IIB)
0.04 mm (Group IIB+H2)
For amepaths not mentioned here, contact AMETEK.

Table of Contents

Related product manuals