EasyManua.ls Logo

Ametek 931S

Ametek 931S
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting | 6-49
Decimal Alarm
Bit Value Type Alarm Condition / Description and Suggested Corrective Action
0 1 Fault f: HC-OBMC communication problem
This serious system Fault condition indicates the Host Controller board cannot
establish reliable communications with the Microcontroller board.
Corrective Action:
Take appropriate safety precautions and cycle the analyzer AC power o󰀨 for a
minimum of 15 seconds before cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
From the Info dialog box and click the HC Reset button to reset the Host
Controller board.
If this still doesn’t correct the alarm, open the Electronics Enclosures, and:
- Reset the Host Controller board (press SW100, Lower Enclosure – see
Figure 6-15).
- Reset the Microcontroller board (press SW400, Upper Enclosure – see
Figure 6-16).
Check the connections of the cable between the Host Controller (J300) and
Micro-Interface (J104) boards. Inspect it for damage (cuts, nicks, burn marks,
etc.).
Replace the Microcontroller, Host Controller, and/or Micro-Interface
boards. Contact AMETEK to verify operation before removing any of these
boards.
1 2 Fault f: Analytical data not received
This serious system Fault condition indicates the Host Controller board is not
receiving analytical data from the Microcontroller board.
Corrective Action:
Take appropriate safety precautions and cycle the analyzer AC power o󰀨 for a
minimum of 15 seconds before cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
From the Info dialog box and click the HC Reset button to reset the Host
Controller board.
If this still doesn’t correct the alarm, open the Electronics Enclosures, and:
- Reset the Host Controller board (press SW100, Lower Enclosure – see
Figure 6-15).
- Reset the Microcontroller board (press SW400, Upper Enclosure – see
Figure 6-16).
Check the connections of the cable between the Host Controller (J300) and
Micro-Interface (J104) boards. Inspect it for damage (cuts, nicks, burn marks,
etc.).
Replace the Microcontroller, Host Controller, and/or Micro-Interface
boards. Contact AMETEK to verify operation before removing any of these
boards.
Setup (tab)Analyzer
Info
Setup (tab)Analyzer
Info

Table of Contents

Related product manuals