EasyManua.ls Logo

Ametek AMP 200N - Page 44

Ametek AMP 200N
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 44 / 60
Test characteristics for CI 250 continuous interference
Method:
Coupling Method : Conducted
Test Regulation Method: Voltage Level (Closed Loop)
Disturbance Current limit : yes/no setted value
Diagram Unit X-Axis : kHz ; Log
Y-Axis: Vpp ; linear
Tolerances
Tolerances : positive +10%
negative -0%
Additional Settings (Closed Loop with Scope only)
Level Change at next Testpoint: -3dB
Couplings
Coupling Type Coupling Network
Coupling Device EM TEST CN 200N
User Software settings
The user has to complete the test settings with his individual used coupling and measuring devices.
New coupling devices can be installed in the menu Setup/Coupling Device by the user.
A: Voltage Measuring Instrument:
Measurement Instrument
Measuring AC voltage ( pk-pk)
Scope
Mandatory in Ford EMC-CS 2009 standard
Instrument Value
Voltage Peak To Peak
B: Measuring Equipment (voltage probe):
Equipment Type
Voltage Probe
Equipment
LeCroy HFP 2500
10:1

Table of Contents

Related product manuals