AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 43 / 60
12.4.5 Immunity to Ground Voltage Offset: CI 250
Components shall be immune from AC ground offset voltages. Requirements include both continuous and
transient disturbances.
Default test setup Continuous Disturbances
Figure 21 Setup for CI 250
RI 250 Test Parameter requirements
Figure 22 Graph for CI 210
Range
Frequency [kHz]
Us [Vp-p]
Frequency step [kHz]
Range 1
2.0 – 10
1.0+1.14 log(f/10)
0.5
Range 2
10 - 100
1.0