AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 51 / 60
Test characteristics for GL VI-7-2-20 (DC) continuous interference
Vector 1:
Name : Vector 1
Start: Level: 3V Frequency 0.05Hz
Stop : Level: 3V Frequency 10.kHz
Step Type: Percentage
Value: 1%
Coupling Method : Conducted
Test Regulation Method: Voltage Level (Closed Loop)
Disturbance Current limit : 1 A or setted value
Diagram Unit X-Axis : kHz ; Log
Y-Axis: V; linear
Tolerances : positive +5%
negative -5%
Additional Settings (Closed Loop with Scope only)
Level Change at next Testpoint: -6dB
Coupling Type Coupling Network
Coupling Device EM TEST CN 200N
User Software settings
The user has to complete the test settings with his individual used coupling and measuring devices.
New coupling devices can be installed in the menu Setup/Coupling Device by the user.
A: Voltage Measuring Instrument:
Measuring AC voltage ( rms)
B: Measuring Equipment (voltage probe):