M470039-01 REV-F  Page  99 
 
MODE CP  or MODE:POW 
This command sets Electronic Load to constant power mode. 
 
MODe:RANGe 
This command sets the range after the mode is set.  
  Parameters:   0 – 3 
i.  low range 
ii.  middle range 
iii.  high range 
iv.  ultra-range  
  Example:   “MODE:RES”  – sets the Electronic Load in CR mode; 
      “MODE:RANG  2” – sets the Electronic Load in high range; 
      “MODE:RANG?  CR” – queries the range for CR mode. 
 
TRANsient     Transient Programming Function 
This subsystem programs the transient functions of three transient modes:  CONTinous, PULSe, TOGGle.  The  input 
levels (TLEVel)  and slew rate (SLEW) are programmed from the current, resistance, power and voltage subsystems. 
 
TRAN[:STATe] 
This command enables or disables the transient or stepping function. The parameter ON | 1 enables the function and OFF 
| 0 disables it. 
  Parameter:  ON | 1, OFF | 0 
  Example:  “TRAN  ON” – enables the transient function 
 
DUTY 
This command sets pulse duty cycle, as a percentage of the total cycle, of continuous transient mode.  If the Electronic 
Load is operating in another mode, the duty cycle is stored until the Load changes to continuous mode. 
  Parameter:  2 to 98; MIN; MAX            Unit: % 
  Example:  “CURR:DUTY 50” – sets a 50% duty cycle for transient current mode. 
 
FREQ 
This command sets pulse frequency of continuous transient mode in Hz.  If the Electronic Load is operating in another 
mode, the pulse frequency is stored until the Load changes to continuous mode. 
   
  Parameter:  0.25 to 20000; MIN; MAX          Unit: Hz 
  Example:  “CURR:FREQ 1000” – sets a 1000Hz frequency for transient current mode. 
 
SLEW 
This command sets the power slew rate for the active mode.  The programmed slew rate is in effect for all changes except 
INP:ON or OFF.  The hardware implementation selects a slew rate that is closest to the SLEW value.  To  determine the 
actual value, use the query SLEW?. 
  Parameter:  0 to maximum value; MIN; MAX        Unit: Depending on mode   
  Example:  “SLEW  6”- sets the slew rate to 6. 
 
TRAN:TWID 
This command sets pulse width of pulse transient mode in millisecond.  If the Electronic Load is operating in another 
mode, the pulse width is stored until the Load is powered cycle.  
   
  Parameter:    0.05 to 10000       Unit: ms 
  Example:    “TRAN:TWID  800” – sets the pulse width to 800 ms.  
  Query syntax:     “TRAN:TWID?” 
  Returned parameter:  Pulse width value in second.  
 
 
STEP     Step Programming Functions 
This subsystem programs the stepping functions for stepping mode.  The number of loops, step values, and dwell times 
are programmed for a specified operating mode.