M470039-01 REV-F Page 107
Use and ” until you see MAIN CC0 parameter. Press enter:
CC0 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CC0 HI = CNT: Input measured voltage from Curr.DMM then press ENTER key. A SUCCESS” message
should be display on the front panel LCD.
Press to select MAIN CC1. Then press enter.
CC1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER
CC1 HI = CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
Press to select MAIN CC2. Then press enter.
CC2 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER
CC2 HI = CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
CONSTANT RESISTANCE (CR0) LOW RANGE MAIN LEVEL CALIBRATION
CR low range main level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of load.
Use and ” until you see MAIN CR0 parameter. Press enter:
CR0 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR0 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CR0 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR0 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
CONSTANT RESISTANCE (CR1) MIDDLE MAIN LEVEL CALIBRATION
CR middle range main level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.
Use and ” until you see MAIN CR1 parameter. Press enter:
CR1 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CR1 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR1 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A SUCCESS” message
should be display on the front panel LCD.
CONSTANT RESISTANCE (CR2) HIGH RANGE MAIN LEVEL CALIBRATION
CR high range main level calibration requires a power supply that has high voltage and medium current, but operator has
to estimate the power consumption of load.