M470039-01 REV-F Page 108
Use and ” until you see MAIN CR2 parameter. Press enter:
CR1 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CR1 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR1 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A SUCCESS” message
should be display on the front panel LCD.
CONSTANT POWER (CP) LOW/MIDDLE/HIGH RANGE, MAIN LEVEL CALIBRATION
CP main level calibration requires a power supply that has enough power output. For example, a PL W6K-60-150 model
needs at least 40V/80A power output to calibrate the unit.
Use and ” until you see MAIN CP0 parameter. Then press enter:
CP0 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP0 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CP0 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP0 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
Use and ” until you see MAIN CP1 parameter. Press enter:
CP1 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CP1 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP1 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
Use and ” until you see MAIN CP2 parameter. Press enter:
CP2 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP2 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CP2 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP2 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
CV TRANCIENT LEVEL CALIBRATION
CV transient level calibration requires a power supply that covers full range of input voltage of load and is capable of
operating in CV and CC mode. Estimate the power consumption of eLoad, because power supply will go to CC mode
while calibrating the load. Adjust the CC current level to a safe limit for load.
Use and ” until you see TRAN CV0 parameter. Then press enter:
CV0 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CV0 HI = CNT: Input measured voltage from Volt.DMM then press ENTER key. A “SUCCESS
should be display on the front panel LCD.
Press to select TRAN CV1. Then press enter.
CV1 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER
CV1 HI = CNT: Input measured voltage from Volt.DMM then press ENTER key. A “SUCCESS message