M470039-01 REV-F Page 109
should be display on the front panel LCD.
Press to select TRAN CV2. Then press enter.
CV2 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER
CV2 HI = CNT: Input measured voltage from Volt.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
CC LOW/MIDDLE/HIGH RANGE, TRANSIENT LEVEL CALIBRATION
CC transient level calibration requires a power supply that covers full range of high range input current of load. Estimate
the power consumption of the electronic load, and adjust the voltage input to a safe limit for the load.
Use and ” until you see TRAN CC0 parameter. Press enter:
CC0 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CC0 HI = CNT: Input measured voltage from Curr.DMM then press ENTER key. A SUCCESS” message
should be display on the front panel LCD.
Press to select MAIN CC1. Then press enter.
CC1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER
CC1 HI = CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
Press to select MAIN CC2. Then press enter.
CC2 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER
CC2 HI = CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
CR LOW RANGE TRANSIENT LEVEL CALIBRATION
CR low range transient level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of electronic load.
Use and ” until you see TRAN CR0 parameter. Press enter:
CR0 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR0 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CR0 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR0 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
CR MIDDLE RANGE TRANSIENT LEVEL CALIBRATION
CR middle range transient level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.
Use and ” until you see TRAN CR1 parameter. Press enter
CR1 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.