M470039-01 REV-F Page 110
CR1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CR1 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR1 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A SUCCESS” message
should be display on the front panel LCD.
CR HIGH RANGE TRANSIENT LEVEL CALIBRATION
CR high range transient level calibration requires a power supply that has high voltage and medium current, but operator
has to estimate the power consumption of load.
Use and until you see TRAN CR2 parameter. Press enter:
CR2 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR2 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CR2 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CR2 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A SUCCESS” message
should be display on the front panel LCD.
CP RANGE TRANSIENT LEVEL CALIBRATION
CP high range transient level calibration requires a power supply that has enough power output. For example, a PLW6K-
60-150 model needs at least 30V/90A power output to calibrate the unit. Only qualified personnel should attempt this
calibration, due to the high level of power.
Use and ” until you see TRAN CP0 parameter. Then press enter:
CP0 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP0 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CP0 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP0 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
Use and ” until you see TRAN CP1 parameter. Press enter:
CP1 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP1 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CP1 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP1 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.
Use and ” until you see TRAN CP2 pa rameter. Press enter:
CP2 LO= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP2 LO= CNT: Input measured voltage from Curr.DMM, then press ENTER.
CP2 HI= CNT: Input measured voltage from Volt.DMM, then press ENTER.
CP2 HI= CNT: Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS message
should be display on the front panel LCD.