105446 06 23-03-2009 117
INPUT SPECIFICATIONS
ACCURACY IN C/F
±0.10°C(±0.17°F) @ 320°C(608°F)
±0.12°C(±0.22°F) @ 650°C(1202°F)
±0.19°C(±0.34°F) @ 1200°C(2192°F)
Accuracy thermocouple type K input
(excluding sensor accuracy)
±0.13°C(±0.24°F) @ -90°C(-130°F)
±0.11°C(±0.20°F) @ -50°C(-58°F)
±0.10°C(±0.17°F) @ 0°C(32°F)
±0.11°C(±0.20°F) @ 155°C(311°F)
±0.12°C(±0.22°F) @ 320°C(608°F)
±0.16°C(±0.28°F) @ 650°C(1202°F)
±0.28°C(±0.50°F) @ 1372°C(2502°F)
Accuracy thermocouple type T input
(excluding sensor accuracy)
±0.14°C(±0.25°F) @ -90°C(-130°F)
±0.12°C(±0.22°F) @ -50°C(-58°F)
±0.10°C(±0.18°F) @ 0°C(32°F)
±0.09°C(±0.16°F) @ 155°C(311°F)
±0.09°C(±0.17°F) @ 320°C(608°F)
±0.10°C(±0.17°F) @ 400°C(752°F)
Accuracy thermocouple type R input
(excluding sensor accuracy)
±1.31°C(±2.35°F) @ -50°C(-58°F)
±0.78°C(±1.40°F) @ 0°C(32°F)
±0.50°C(±0.90°F) @ 155°C(311°F)
±0.42°C(±0.75°F) @ 320°C(608°F)
±0.41°C(±0.74°F) @ 650°C(1202°F)
±0.50°C(±0.90°F) @ 1760°C(3200°F)
Accuracy thermocouple type S input
(excluding sensor accuracy)
±0.98°C(±1.77°F) @ -50°C(-58°F)
±0.78°C(±1.40°F) @ 0°C(32°F)
±0.50°C(±0.90°F) @ 155°C(311°F)
±0.46°C(±0.83°F) @ 320°C(608°F)
±0.45°C(±0.81°F) @ 650°C(1202°F)
±0.52°C(±0.94°F) @ 1768°C(3214°F)
Accuracy thermocouple type B
(excluding sensor accuracy)
±1.57°C(±2.83°F) @ 250°C(482°F)
±0.99°C(±1.78°F) @ 320°C(608°F)
±0.69°C(±1.23°F) @ 650°C(1202°F)