THERMAL SPECIFICATIONS –
DRY - BLOCK
ATC-140 A/B
Reference accuracy ±0.02°C / ( ±0.04°F)
Total accuracy ±0.18°C / (±0.32°F)
Heating time incl. insert -20°C / -4°F to 23°C / 73.4°F : 10 min.
23°C / 73.4°F to 100°C / 212°F: 31 min.
100°C / 212°F to 140°C / 284°F: 23 min.
23°C / 73.4°F to 140°C / 284°F : 54 min.
-20°C / -4°F to 140°C / 284°F : 64 min.
Time to stability 15 min.
Cooling time incl. insert 140°C / 284°F to 23°C / 73.4°F : 34 min
140°C / 284°F to 100°C / 212°F : 7 min.
100°C / 212°F to 23°C / 73.4°F: 27 min.
23°C / 73.4°F to -15°C / 5°F : 52 min.
23°C / 73.4°F to -20°C / -4°F: 95 min.
23°C / 73.4°F to 0°C / 32°F: 17 min.
0°C / 32°F to –15°C / 5°F: 35 min.
0°C / 32°F to –20°C / -4°F : 78 min.
140°C / 284°F to –20°C / -4°F : 129 min.
THERMAL SPECIFICATIONS –
LIQUID BATHS
ATC-140 A/B
Maximum temperature 140°C / 284°F
Minimum temperature
-33°C / -27°F @ ambient temperature 0°C / 32°F
-18°C / 0°F @ ambient temperature 23°C / 73.4°F
-3°C / -26°F @ ambient temperature 40°C / 104°F

Table of Contents

Other manuals for Ametek JOFRA ATC-125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ATC-125 and is the answer not in the manual?

Ametek JOFRA ATC-125 Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ATC-125
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals