9
OPERATION MANUAL
© AMETEK, Inc.
4 FEATURES MINIFLASH FP(H) VISION
Only 1 or 2 ml (depending on measuring standard) of sample for a measurement
No open flame
No offensive fumes
Measurement is fully automatic
Accurate and reliable determination of flash points
Fully automatic barometric pressure correction
Safe operation due to completely closed cup (Ignition Protection Technology)
No accessories required
Wide temperature range:
FP VISION: -25°C to 120°C (-13°F to 248°F) with external cooling
FPH VISION: 10°C to 400°C (50°F to 752°F)
Built-in diagnostic and safety features
Lightweight and compact, portable
Power from car battery via power converter
Intuitive Menu Navigation on large 10” Color Touch-Screen
Runs on Linux software
Interfaces:
4x USB for data transfer, printer, mouse, keyboard and BARCODE-Reader connection
2x LAN for full network, PC, LIMS and remote operation
2x HDMI
USB printer support and data transfer
COCKPIT interface
Combustion Analysis, Peltier Protection Technology
64GB Harddisc for result storage
Automatic Ignition Cleaning Program
Multilingual display
4.1 Technical data
Instrument Data
°
°
Communication languages Multilingual (English, German, French, et. al.).
Power requirements 100/110/120/230/240 V AC; 50/60 Hz;
Environmental Certificates EN 60068-2-1, EN 60068-2-78, EN 60068-2-14
Shockproof EN 60068-2-6 (Requ. IEC 60721-3-2, Class 2M2)
Optional DC / AC Power converter for use with car battery
Fuse Type FST 5 x 20 mm time lag T2A/250V, IR35A/250V,
AC IEC 60127-2/3
Temperature range
FP VISION -25°C to 120°C (-13°F to 248°F)
with external cooling
FPH VISION 10°C to 400°C (50°F to 752°F)
Accuracy of temperature reading
°
°
Pressure sensor accuracy
0.5 kPa ( 0.073psi)
Physical dimensions W x H x D = 253 x 368 x 277 mm
W x H x D = 10 x 14.5 x 10.9 inches
Environmental Specifications
Operating Temperature: 0°C to 50°C (32°F to 122°F)
Operational humidity: ≤ 90% RH non condensing
Weight FP(H) VISION 11 kg (26 pounds)