34
© AMETEK, Inc.
8 CREATE A CUSTOM MEASUREMENT
1 … In the Main Menu choose at Committee
custom to create a new custom program
2 … Then the parameter section is extended and
the whole parameters appear
Ti … Initial temperature where the
measurement starts. The oven will be preheated
to this temperature before the cup is lifted.
Tf … Final temperature where the measurement
cycle stops. In case a flashpoint is detected, this
temperature won’t be reached. In case no
flashpoint is detected, the measurement is
stopped with the result No flash.
Ti-T … this is the parameter to cool the sample
in case the measurement has to start below
ambient temperature.