55
MINIFLASH
FP(H) VISION
OPERATION MANUAL
© AMETEK, Inc.
15.3 Clean temperature probe
In order to get accurate temperature readings for tests the temperature probe must be cleaned regularly. Therefore,
use the brass brush to free the temperature probe from residuals.
When doing so take care not to bend the temperature sensor (2) or the arc pins (1).
15.4 Cleaning of the sample cup
In order to get accurate flashpoints for tests the sample cup must be cleaned properly after each test. Therefore, use
disposable tissues in combination with a cleaning agent to clean the cups from its sample residuals.
Take care that cleaning agents are evaporated already prior filling with fresh sample.
If cleaning of a sample cup is not done properly, wrong flashpoints are detected due to contaminations of previous
tested samples.