15 PREVENTIVE MAINTENANCE
15.1 Clean heater
The surface of the heater must be free of residuals and scratches for good heat contact.
Use a tissue in order to clean the surface of the heater.
When doing so take care not to bend the temperature sensor (2) or the arc pins (1).
15.2 Clean arc pins
In order to get accurate ignitions for tests the arc pins must be cleaned regularly. Therefore use the brass brush to free
the arc pins from soot or other residuals.
When doing so take care not to bend the temperature sensor (2) or the arc pins (1).