Table 9 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
Immunity to Proximity Fields from RF Wireless Communications Equipment
Guidance and Manufacturers Declaration - Electronic Immunity


Immunity
Test
IEC 60601Test Level
Compliance
Level
Electromagnetic
Environment
-Guidance-
Radiated RF
IEC 61000-4-3
Test
Frequency

  Modulation
Maximum

Distance

Immunity
Test Level

Compliance
Level

where
d = Minimum
separation
distance in
meters
E = Immunity
test level in

P = Maximum
power in

385 380-390 TETRA 400
Pulse Modulation
18 Hz
1,8 0,3 27

0,3 m
450 430-470
GMR 460, FRS
460
FM ±5 kHz devia-
tion 1 kHs sine
2 0,3 28

0,3 m
710
704-787 
Pulse Modulation
217 Hz
0,2 0,3 9 745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820, CDMA

3, 4, 25; UMTS
Pulse Modulation
18 Hz
2 0,3 28

0,3 m
870
930
1720
1700-1990
GSM 1800; CD-
MA 1900; GSM
1900; DECT; LTE

UMTS
Pulse Modulation
217 Hz
2 0,3 28

0,3 m
1845
1970
2450 2400-2570

802.11 b/g/n,
RFID 2450, LTE

Pulse Modulation
217 Hz
2 0,3 28

0,3 m
5240
5100-5800 WLAN 802.11 a/n
Pulse Modulation
217 Hz
0,2 0,3 9 5500
5785