EasyManuals Logo

Ametek VDS 200Q Series User Manual

Ametek VDS 200Q Series
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 22 / 36
3.5.
Source Settings
The source settings menu contains features that are specific to the VDS 200Q series.
SOURCE SETTINGS
Vmax Pos
= 80.0V
Vmax neg
= -20.0V
Gain
= 8x
I peak
= NO Peak
Freq mode
= HF
Impedance
= 10 mR
Vnom
= 13.5V
I
= 200 A
DEFAULT
Vmax Pos
Sets the limit of the maximum positive voltage that can be used. This is to protect DUTs and sensitive setups.
Vmax Neg
Sets the limit of the maximum negative voltage that can be used. This is to protect DUTs and sensitive setups.
Gain
The VDS 200Q has two gain settings. The normal operation is 8, but the user can also select a gain of 4. Note,
however, that with a gain of 4, the maximum voltage output of the amplifier becomes 40V. For tests that do not
require over 40V, gain 4 is recommended because the VDS 200Q is also much more efficient.
I peak
The current limitation of the VDS 200Q is very advanced. There are three settings:
3x Imax: This selection will allow an inrush current of three times the maximum current available from the
generator for 200 ms before the current limitation starts. A VDS 200Q200 can produce 600A for
200ms, after which the voltage will be regulated down to achieve the maximum current limitation set.
3x Ictrl: This selection will allow an inrush current of three times the programmed current limit for 200 ms
before the current limitation starts.
Peak OFF: This selection enables no inrush current above what is set. The current limit circuitry will start
immediately.
Frequency mode
The VDS 200Qxx.2 supports three frequency ranges,
Std: Standard. This is the normal operating mode with a good compromise between stability and
bandwidth. The bandwidth in this selection is approximately 40 kHz.
Capa: Capacitive. This mode is specifically designed for capacitive and/or reactive loads. This reduced
bandwidth, high stability mode is perfect for use during normal transient testing, conducted sine wave
(CSW) testing. The bandwidth in this mode is limited to approximately 3 kHz. This is the default
startup mode, and is still fast enough for most common pulses like Pulse 4 and 2b.
HF: This mode is for testing where high bandwidths are required. This mode allows frequencies up to 250
kHz.
CAUTION:
As with all fast amplifiers, the user should monitor the DUT to ensure no ringing or
excessive over- shoot is present that could result in damage to the DUT or couplers. In case
there is a question the Capacitive mode should be used.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek VDS 200Q Series and is the answer not in the manual?

Ametek VDS 200Q Series Specifications

General IconGeneral
BrandAmetek
ModelVDS 200Q Series
CategoryAmplifier
LanguageEnglish

Related product manuals