5.10 Adjuvant treatment
Note:
[Adjuvant treatment] Using other medical devices or drugs with E6 will
affect the treatment. For other devices and/or drugs, be sure to refer to
their Instructions for Use to ensure they are suitable for use with CPR.
5.10.1 Defibrillation
Defibrillation can be performed while the E6 is running.
1. You can use the defibrillation electrodes before or after positioning the E6.
2. Defibrillation according to the instructions of the defibrillator
manufacturer.
Note:
[Defibrillation Electrodes] Position the defibrillator electrodes and wires so
that they are not under the suction cup. If the electrode already exists on the
patient, make sure it is not under the suction cup. If it is under the suction
cup, a new electrode must be used.
3. After the defibrillation is finished, make sure the position of the suction
cup is correct. Adjust the position if necessary.
Warning:
[Position changed during operation] If the position of the suction cup
changes during operation or defibrillation, immediately press the adjustment
button to adjust the position. Always use the E6 stabilizing beltp to help
ensure the correct position.
[ECG interference] Chest compression can interfere with ECG analysis.
Press Pause first before starting the ECG analysis. Reduce interruption time
as much as possible. Press Start (continuous) or Start (15:2, 30:2) to restart
pressing.
5.10.2 Ventilation
Always follow local and / or international ventilation guidelines.
E6 can be run in two different modes:
Start (continuous)
When you select this mode, the E6 is continuously pressing and, the
ventilation is controlled by the operator or other breathing apparatus.