Fig. 38 Start
5.6.4 Move patient
E6 can be enabled when you move a patient under the following conditions:
• E6 is safe with the patient on a transport unit.
• E6 maintains the correct position and angle on the patient's chest.
Adjust the position of the suction cup if necessary.
Note:
• [Position changed during operation] If the position of the suction cup
changes during operation or defibrillation, immediately press the
adjustment button to adjust the position. Always use the E6 stabilizing
belt to help ensure the correct position.
5.7 Change battery during operation
When the battery is about to run out, E6 sends an alarm by lighting an alarm
light (red alarm), displaying an alarm message in the message bar, and an
audible (or voice) alarm.
5.7.1 Change battery
Note : To minimize disruption, we offer a dual battery design. It is
recommended to replace the new battery or connect the external power
supply to charge the battery after a battery has a low battery alarm. Try to
avoid the situation where only one battery is left.
1. When the E6 is connected to an external power source or there is still a